    The Calibration Kit has been created to provide all

    the necessary components to enable customers to

    verify the accuracy of the readings on the Buckleys or

    Roxby Bathycorrometers so that in-house

    calibration certificates can be issued.

    The process requires two areas to be checked. The

    electronics should be tested using the BCM Checker.

    Readings on the Bathycorrometer should be 0.000

    with the checker switched off and 1.990 with the

    checker on.

    Providing the readings are correct, the Silver/Silver

    Chloride screw-in half cell should be checked by

    using K-Series Reference Cells in a standard salt

    water solution. To confirm the readings, three cells

    should be used and the Sea Water corrosion test

    mixture should be made up to a 3% solution.